3. Add, delete, or change commands as necessary, then click Save As.
NOTE: When you choose the CAL command, a Rinse command is
performed prior to the CAL. When the CAL is complete, two
additional Rinse commands are performed. This prevents a bubble
from loading into the chamber. Verify that you have sufficient room in
the reservoir to complete these Rinse commands.
When you choose the VER command, a Rinse command is
performed prior to the VER. When the VER is complete, an
additional Rinse command is performed. This prevents a bubble
from loading into the chamber. Verify that you have sufficient room in
the reservoir to complete these Rinse
commands.
4. Type a different name for the routine to create a new routine, or type the existing name of
the routine to keep the edited routine using the existing routine name.
NOTE: You can only add commands to the end of a routine. You cannot
insert new commands before commands that are already part of a
routine.
Deleting a Routine
You can delete a routine that you created, but you cannot delete predefined routines.
Predefined routines have (Luminex) after the routine name.
1. Open the Maintenance
page, then open the Cmds & Routines tab.
2. Click the routine that you want to delete from the Routine Name list.
3. Click Delete.
Running a Routine
1. Open the Maintenance page, then open the Cmds & Routines tab.
2. Select a routine to run from the Routine Name list.
3. Click Eject on the status bar.
4. Add the appropriate reagents to the plate, reservoirs, and well strips as indicated in the
plate image and set the plate on the plate holder.
5. Click Retract.
6. Click Run. The Routine Message dialog box opens when the routine is complete.
7. Click OK.
Importing a Routine
1. Open the Maintenance page, then open the Cmds & Routines tab.
2. Click Import.
3. In the Open dialog box, browse for the file you want to import, then click Open. The
routine becomes the active routine.