Minimum bead count for obtaining results - Type any whole number from 0 to 1000000.
This is the minimum bead count that must be observed by the instrument for a particular
analyte before this data is used in statistical calculations and is displayed on the graphs
and results table. This number varies based on assay. When set to a number greater than
0, the analyzer does not display data for bead sets not generating events equal to or less
than this value. Type 0 to display all events. The default setting is 1.
Allow batches to be run or saved without lot number, expiration or manufacturer -
Allows saving lots without normal required information.
Run Routine after batch complete - Select this option to specify a particular routine to
run after a batch is finished. If you select this option, then select a routine from the Routine
to run after batch completes list.
Run Rinse before batch starts - This options performs a Rinse at the last selected
location on the Cmds and Routines page for the Rinse command. Ensure that you have
sufficient room in the reservoir to perform this Rinse.
Batch Thresholds
Low bead count detection - Selecting either Error
or Warning stops the batch or logs a
warning if the number of successive wells in the Well Count box are run without reaching
the number of beads in the Bead Count box. Click in either box to type a number other
than the default. Warning pause pauses the batch to allow the user to change the load
volume and then resume the batch.
Cancel - Cancels the changes.
Save - Saves the changes.
Alert Options Tab
To view this tab, click the Alert Options tab on the Admin page. Use this tab to set options
for alerts about various system events.
Application Administrator Tasks
21