Luminex xPONENT for MAGPIX User Manual Page 64

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 63
1. Open the Batches page.
2. In the Pending Batches
section, click the batch you want to delete, then click Delete.
The Delete Pending Batch dialog box opens.
3. Click Yes.
NOTE: Batches saved to a multi-batch cannot be edited or deleted unless
they are removed from the multi-batch. However, you can edit the
multi-batch itself. To remove a batch from a multi-batch, click a well
in the plate layout, then click Remove.
NOTE: To delete a partial batch, open the Admin page, then click the
Archive tab. Clear the Keep data after archive box. Click Archive
to delete the partial batch.
Editing a Batch
1.
Open the
Batches
page.
2. Click the batch you want to edit, then click Edit. The Protocol tab opens.
3. Edit the information as needed on the Protocol, Std & Ctrls, and Plate Layout tabs. For
the Plate Layout tab, confirm that the plate layout conforms to your specific assay
instructions.
4. Click Save on the Plate Layout tab.
NOTE: Batches saved to a multi-batch cannot be edited or deleted unless
they are removed from the multi-batch. However, you can edit the
multi-batch itself. To remove a batch from a multi-batch, click on a
well in the plate layout, and click Remove.
Settings Tab
Use this tab to name your new batch and configure acquisition settings. For existing batches,
you can view the acquisition parameters of the selected saved batch and print the batch
settings report.
Using xPONENT
51