Cleaning the Ventilation Filters
Each filter has at least one imprinted arrow on its metal frame. This arrow indicates the
direction of the air flow. Filters should be installed so that the arrows are at the correct
locations and point in the correct direction.
To clean the Luminex
®
FLEXMAP 3D
®
ventilation filters:
1.
Turn off the FLEXMAP 3D
®
instrument and unplug the power cord.
2. Remove the XY cover by gently grasping its ventilation holes and pulling it straight out.
3. Slide clamps off the silver filter inside the cover and remove the filter.
NOTE: For filter location, reference the "Ventilation Filter on XY Platform"
image below.
4. Open the front doors. Slide clamps off the left and right silver door filters and remove the
filters.
NOTE: For clamp location, reference the "Ventilation Filters inside Doors"
image below.
5. Clean the filters with a vacuum or with deionized water.
6.
Stand the filters upright to air dry.
CAUTION: Filters must be completely dry prior to reinstallation.
7. Reinstall filters. The arrows on the door filters should point to the inside of the door. The
arrow on the XY filter should point up.
For research use only. Not for use in diagnostic procedures.
Maintenance and Cleaning