6. Connect the waste fluid tubing (orange connector) to the bottom quick disconnect point
(white connection) of the sheath filter tubing.
7.
Use the software to run a warmup command. This starts the compressor, which drains
the sheath fluid from the reservoir into the waste fluid container.
NOTE: To refill the reservoir, reconnect each connector to its color-matched
connection and use the software to run a prime command.
FIGURE 23.
Front of Fluidics Bay
1 White connection for sheath fluid tubing - Disconnect as referenced in Step 3
2 Orange connection for waste fluid tubing - Disconnect as referenced in Step 4
3 Blue connection for sheath filter tubing - Disconnect as referenced in Step 5
Shipment Checklist
Complete the following checklist, sign and date it, and return it with the
Luminex
®
FLEXMAP
3D
®
system.
1. Remove all specimens from the instrument.
2. Decontaminate the instrument. See the section titled “Decontamination Procedure”.
3. Drain the reservoir. See the section titled “Draining the Reservoir”.
Was there an internal leak in the system?
Luminex
®
FLEXMAP 3D
®
Hardware User Manual
54
For research use only. Not for use
in diagnostic procedures.