Symptom Possible cause Solution
The microspheres are photobleached. Replace the microspheres with
unphotobleached microspheres.
There are an insufficient number of beads
in the sample.
Ensure that there are 2000 to 5000 beads
per bead set per well.
The sample is too concentrated. Dilute concentrated biological fluids, such
as serum or plasma, at least 1:5.
Bead Detail Irregularities
Use these tools to assist in diagnosing system and kit-related problems:
instrument calibrators
instrument verifiers
assay standards
assay controls
error messages
Review calibration/control trend reports routinely to detect trends.
Use instrument Luminex
®
xMAP
®
verification microspheres to check the success of the
instrument calibration and to troubleshoot. If there is a problem with kit results, xMAP verifiers
can help determine if the problem is related to the instrument. If calibration and verification
are successful, contact the kit manufacturer.
The Bead Detail Irregularities table displays irregular groupings of beads. Compare the
displays to the following bead detail of a normal bead grouping - a tight bead population
within a white region.
FIGURE 22.
Normal Bead Detail
TABLE 8.
Verification Problems Continued
Luminex
®
FLEXMAP 3D
®
Hardware User Manual
50
For research use only. Not for use
in diagnostic procedures.