2. Remove the sample probe from the instrument and flush it with deionized water from the
narrow end out through the larger end.
3.
Replace the sample probe in the sample arm and wrap the end of the probe with
Parafilm
®
.
Preparing the Luminex
®
FLEXMAP 3D
®
Instrument for Use
After Storage
Follow this procedure before you begin using the Luminex
®
FLEXMAP 3D
®
instrument after
its removal from long-term storage.
To prepare the FLEXMAP 3D
®
instrument for use after storage:
1. Make sure that the sheath container has a sufficient amount of sheath fluid and that the
waste fluid container is empty.
2. Remove the Parafilm from the end of the sample probe.
3. Turn on the FLEXMAP 3D
®
instrument and watch for the following indications of correct
response:
The compressor starts. It has a low rumbling sound.
Air blows out of the rear fans. Place your hand behind the FLEXMAP 3D
®
instrument,
on the upper part of the back, to feel it.
NOTE: For the fan location, see the "Back of the FLEXMAP 3D
®
Instrument"
image below.
The syringes inside the right door of the FLEXMAP 3D
®
instrument initialize.
4. Turn on the PC and start up the software. Use the software to run a revival after storage
routine.
Replacing Fuses
Periodically, you may need to replace a fuse on the Luminex
®
FLEXMAP 3D
®
instrument. The
instrument requires fuses with the following specifications:
F6A, 250 V
The fuse cartridge will accept either 5 mm x 20 mm or 0.25” x 1.25” fuses. Fuses are
available from Luminex Corporation. Replacing a fuse requires access to the back of the
FLEXMAP 3D
®
instrument. Be aware before attempting any necessary lifting or movement of
the instrument that it weighs approximately 91 kg (200 lbs). Lifting it requires five people.
DANGER: To avoid serious injury or death by electric shock, you must turn
off the FLEXMAP 3D
®
instrument and unplug it from the wall
before replacing a fuse.
To replace a fuse:
1.
Unplug the power cable from the instrument.
2. Use a small, flat-blade screwdriver to open the module door on the lower left corner of
the back of the instrument. The door opens downward. Inside are two cartridges, a red
one on top and a black one on the bottom.
3. Use the screwdriver to remove the red cartridge.
For research use only. Not for use in diagnostic procedures.
Maintenance and Cleaning