TABLE 9.
Bead Detail Irregularities
Symptom Description Possible Problem Solution
xMAP and calibrator microspheres form a group in the wrong location.
Calibrator microspheres group
together at the top of the
region.
The calibrator microspheres
are photobleached.
Replace the calibrator microspheres with
unphotobleached calibrator microspheres.
To avoid photobleaching, protect
microspheres from light.
xMAP microspheres group
together at the lower right of
the region.
The xMAP microspheres are
photobleached.
Replace the xMAP microspheres with
unphotobleached xMAP microspheres. To
avoid photobleaching, protect microspheres
from light.
xMAP microspheres form an unusual shape.
xMAP microspheres form a
long diagonal line.
The xMAP microspheres have
agglutinated.
Contact Technical Support.
xMAP microspheres form a
group with extending tails.
The solvent is incompatible. The Luminex Technical Support website
(http://www.luminexcorp.com/Support) lists
incompatible solvents. If the solvent you are
using is listed, switch solvents.
xMAP microspheres fail to form a group.
xMAP microspheres do not
form a cohestive group.
The sheath fluid is
incompatible.
Use only Luminex xMAP sheath fluid in the
Luminex
®
FLEXMAP 3D
®
instrument. Other
fluids can damage the instrument and void
your warranty.
For research use only. Not for use in diagnostic procedures.