View A Right front door, viewed from inside 2 Filter Clamp
View B Left front door, viewed from inside 3 Incised Arrow
1 Filter
Annual Maintenance
Annual maintenance requires replacing the sheath filter.
To replace the Luminex
®
FLEXMAP 3D
®
instrument sheath filter:
1. Turn off the FLEXMAP 3D
®
instrument and unplug the power cord.
2. Open the left door on the FLEXMAP 3D
®
instrument and locate the sheath filter on the
front of the fluidics bay.
3. Disconnect the filter by pushing down on the metal clamps at each quick-disconnect
point.
NOTE: For information on the location of the disconnection clamps,
reference the "Front of Fluidics Bay (Inside Left Door)" image below.
4. Connect the new sheath filter, matching up the color-coded fittings. The arrow on the
sheath filter should be pointing up.
5.
Close the left door.
For research use only. Not for use in diagnostic procedures.
Maintenance and Cleaning
37